SERVICING AND CLE ANING
Mount piston
6.
Reassemble piston
7.
and nose cone
assembly
Mount maifold
8.
Performe
9.
reference run
Insert the cleaned or new piston. Screw on the
piston nut and tighten with the mounting tool.
Loosen the nose cones by a half rotation.
Afterward push the nose cone assembly on
the piston unit and fasten it. Then tighten nose
cones.
Reassemble the manifold and mount it to the
handle.
NOTE:
The manifold must be screwed counter-clock-
wise into the handle so that it audibly snaps
into place.
You must not pull the manifold downward.
Check the instrument for tightness, mobility
and accuracy (see page 27 "Servicing").
Performe the reference run (rEF).
ErgoOne E-8/-12 · User Manual · 31