SERVICING AND CLE ANING
Remove seal
4.
Mount seal
5.
Reassemble
6.
manifold
Perform
7.
reference run
Push the piston unit down to the bottom. Re-
move the seal, inspect it and clean or replace if
necessary.
NOTE:
The seal will either remain inside the nose cone
or will stay on the piston after the nose cone is
removed.
If required, grease the piston lightly with the
supplied silicone grease. Push the seal on the
piston with its flat side facing upward. Use
the mounting tool to mount and tighten the
cleaned or new nose cone.
Reassemble the manifold and mount it to the
handle.
NOTE:
The manifold must be screwed into the handle
counter-clockwise, until it audibly snaps into
place.
You must not pull the manifold downward.
Check the instrument for tightness, mobility
and accuracy (see page 27 "Servicing").
Perform the reference run (rEF).
ErgoOne E-8/-12 · User Manual · 29