SERVICING AND CLE ANING
Removing of pistons for cleaning or replacing
Unscrew manifold
1.
Slide housing
2.
Remove screws
3.
Separate piston
4.
and nose cone
assembly
Remove piston
5.
30 · ErgoOne E-8/-12 · User Manual
Unscrew the manifold from the handle
(See p. 28).
Turn both closures of the manifold housing
cover 90° and slide off the housing.
Remove both outer Phillips screws on the nose
cone assembly.
ATTENTION:
Do not loosen the central guide rod (Z)!
Pull the piston and nose cone assembly apart
and separate.
Place the mounting tool on the piston nut and
unscrew the piston nut. Remove the piston nut
and pull out the piston.