E A S Y C ALIB R ATION ADJUS TMENT OF THE ERGOONE® E-8/-12
The calibration mode 'CAL'
Adjustment
The instrument should be set to either the nominal volume (for example
100 µl for a 100 µl pipette) or a specific test volume, in the standard
pipetting mode (PIP). See page 8 and 12 for procedures.
E. g., volume according to testing of volume 101.3 µl.
What to do
Bring up the CAL
1.
mode
Activate the CAL
2.
mode
Confirm CAL mode
3.
Set the volume
4.
Confirm volume
5.
24 · ErgoOne E-8/-12 · User Manual
How to do it
Press and hold the MENU
key (> 3 sec) until CAL
appears. The display reads
'off'. 'CAL' blinks.
Press either the UP or
DOWN key to activate the
CAL mode. The display
changes from 'off' to 'on'.
'CAL' continues to blink.
Press the ENTER key. The
display now shows the set
pipetting volume.
'CAL' blinks.
Use the arrow keys (+/-) to
set the volume, which was
previously determined and
tested. 'CAL' blinks.
Press the ENTER key.
The display shows the
tested and corrected
volume. The CAL symbol
is continously displayed to
confirm that an adjustment
has been made.
Keys to press
1x
>3s
1x
1x
Display readout