ACCUR ACY TABLE
Final test values related to the nominal capacity (maximum volume)
and the indicated volume steps indicated on the instrument, obtained
when instrument and distilled water are equilibrated at ambient tem-
perature (20 °C/68 °F) and with smooth operation. According to
DIN EN ISO 8655.
Accuracy tolerances for the ErgoOne
Volume range
µl
0.5 -
10
1 -
20
5 - 100
15 - 300
* A = Accuracy, CV = Coefficient of Variation
E-8/-12
®
Volume step
A*
<
µl
+ %
10
1.2
5
2.0
1
8.0
20
1.0
10
2.0
2
8.0
100
0.8
50
1.6
10
4.0
300
0.6
150
1.2
30
3.0
CE
CV*
Increment
<
%
µl
0.8
0.01
1.5
4.0
0.5
0.02
1.0
3.0
0.25
0.1
0.4
1.5
0.25
0.5
0.4
1.2
ErgoOne E-8/-12 · User Manual · 23
20 °C
!
Ex
Recommended
Type of tip, µl
20
20
200/300
200/300