AUTOCL AVING
The manifold (M) of the ErgoOne
is autoclavable at 121 °C (250 °F), 2 bar absolute (30 psi) with a holding time
of at least 15 minutes, according to DIN EN 285.
ATTENTION:
1.
Eject the pipette tips.
2.
Unscrew the manifold from the grip (see page 28).
3.
Autoclave the complete manifold without any further
disassembling.
4.
Allow the manifold to completely cool and dry.
5.
Screw the manifold into the grip again (see page 28).
6.
Perform a reference run (rEF).
NOTE:
The effectiveness of the autoclaving must be verified
by the user.
Maximum reliability is obtained with vacuum steriliza-
tion. We recommend the use of sterilization bags.
If the manifold is autoclaved frequently, then the pistons should
be greased with the supplied silicone grease in order to provide
smoother movement.

REFERENCE RUN (REF)

A manual reference run must be completed each time the manifold is reattached
to the handle. The reference run is needed to assure secure connection of the piston.
What to do
Bring up rEF mode
1.
Perform the
2.
reference run
NOTE:
After the reference run, the display automatically returns to the previous program.
26 · ErgoOne E-8/-12 · User Manual
E-8/-12 (highlighted in picture)
®
The handgrip can not be autoclaved!
How to do it
Simultaneously press the
MENU and the ENTER key
to activate the rEF mode.
Press the pipetting key
once to start the reference
run. A noise can be heard,
clearly indicating the func-
tion is being performed.