E A S Y C ALIBR ATION
Revert to factory default settings
The continually displayed CAL symbol refers to a previously made
adjustment.
What to do
Bring up the CAL
1.
mode
Deactivate CAL
2.
mode
Revert to factory
3.
setting
When the ErgoOne
IMPORTANT:
performed, which means that the volume is changed across
the entire volume range of the pipette by the same amount.
It is recommend that the adjustment be performed at 50% of
the nominal volume.
The instrument is permanently adjusted for watery solutions,
NOTE:
but it can also be set for solutions with varying density, viscos-
ity and temperature. The ErgoOne
every mode, with the exception of the GEL mode.
How to do it
Press and hold the MENU
key (> 3 sec) until CAL
appears. The display reads
'on'. 'CAL' blinks.
Press either the UP or
DOWN key to deactivate
the CAL mode. The display
changes from 'on' to 'off'.
'CAL' continues to blink.
Press the ENTER key.
The CAL symbol disap-
pears. The instrument has
now been reverted to fac-
tory default setting.
E-8/-12 is adjusted, a volume offset is
®
Keys to press
1x
>3s
1x
E-8/-12 can be adjusted in
®
ErgoOne E-8/-12 · User Manual · 25