Detailed information about your system
Precision
Within Run Precision (300 Venous Blood Samples Tested per
Glucose Level)
Data generated using the OneTouch Verio Flex
Target
Glucose
(mmol/L)
2.2
5.0
7.2
11.1
19.4
Results show that the greatest variability observed between
test strips when tested with blood is 0.10 mmol/L SD or less
at glucose levels less than 5.55 mmol/L, or 1.98% CV or less
at glucose levels at 5.55 mmol/L or above.
Mean
Glucose
(mmol/L)
2.0
4.9
7.1
11.1
19.1
87
®
Standard
Deviation
(mmol/L)
0.05
0.10
0.14
0.22
0.34
9
Meter.
Coefficient
of Variation
(%)
2.26
1.95
1.95
1.98
1.79