Detailed information about your system
Reported result range
Sample
Sample volume
Size
Test time
Unit of measure
Weight
System Accuracy
Diabetes experts have suggested that 95% of glucose meter
results should agree within 0.83 mmol/L of a laboratory
method when the glucose concentration is lower than
5.55 mmol/L, and within 15% of a laboratory method when
the glucose concentration is 5.55 mmol/L or higher.
Samples from 100 patients were tested using both the
OneTouch Verio Flex
Analyzer laboratory instrument.
1.1-33.3 mmol/L
Fresh capillary whole blood
0.4 µL
52.0(W) x 86.0(L) x 16.0(T) mm
5 seconds
mmol/L
Approximately 50 grams
System and the YSI 2300 Glucose
®
83
9