Detailed information about your system
9
System Accuracy Results for Glucose
Concentrations <5.55 mmol/L
Percent (and number) of meter results that match the
laboratory test
Within
±0.28 mmol/L
82.0% (123/150)
System Accuracy Results for Glucose
Concentrations ≥5.55 mmol/L
Percent (and number) of meter results that match the
laboratory test
Within ±5%
62.9% (283/450)
Within
±0.56 mmol/L
98.0% (147/150)
Within ±10%
94.7% (426/450)
84
Within
±0.83 mmol/L
100% (150/150)
Within ±15%
99.3% (447/450)