Detailed information about your system
9
Regression Statistics
Samples were tested in duplicate on each of three test strip
lots. Results indicate that the OneTouch Verio Flex
compares well with a laboratory method.
# of Subjects
100
95% CI Slope
1.00 to 1.01
User Performance Accuracy
A study evaluating glucose values from fingertip capillary
blood samples obtained by 167 lay persons showed the
following results:
93.1% within ±0.83 mmol/L of the medical laboratory values
at glucose concentrations below 5.55 mmol/L, and 98.6%
within ±15% of the medical laboratory values at glucose
concentrations at or above 5.55 mmol/L.
97.6% of the total number of samples were within
±0.83 mmol/L or ±15% of the medical laboratory values.
# of Tests
600
95% CI
Intercept
(mmol/L)
-0.12 to 0.08
Slope
1.01
Std. Error
(S
)
y.x
(mmol/L)
0.64
86
System
®
Intercept
(mmol/L)
-0.02
R
2
0.99