Télécharger Imprimer la page

Leica Microsystems DM2000 Manuel D'utilisation page 134

Publicité

Les langues disponibles

Les langues disponibles

9. Contrast Methods
Magnifications of 1.6x and 2.5x
are also possible with the CL/PH or CLP/PH, UCL
or UCLP condensers if the condenser is re-
moved completely. The field diaphragm then
takes over the function of the aperture dia-
phragm.
Note:
If the microscope is equipped for polarization,
the analyzer and polarizer as well as the lambda
plate compensator must be removed or swung
out when using other contrast methods.
9.1.1 Brightfield
• If present: click the condenser disk* into the
BF position.
• If present: pull the light ring slide* out of the
condenser.
• If present: switch the fluorescence illuminator
into an empty position or filter system A.
• Insert a transmitted light specimen.
• Rotate an appropriate objective into place.
• Movable condenser heads:
The condenser top is swung out for objec-
tive magnifications < 10x.
• Bring the image into focus using the focus dial
and set the brightness.
• For an optimal field diaphragm setting, check
the Köhler illumination (→ p. 34).
• Use suitable transmitted light filters as appli-
cable (Fig. 60).
56
Fig. 60
Filter holders
Leica DM2000/3000 only:
DLF filter magazine for
attachment to micro-
scope base
Leica DM2000/3000 only:
Filter holder with 2 posi-
tions or 1 position for
attachment to micro-
scope base
Filter holder for screw
attachment
on
con-
denser
Leica DM2500 only:
Adapter
with
filter
holders
between
stand and LH 107/2

Publicité

loading

Ce manuel est également adapté pour:

Dm2500Dm3000