6.6.3 Equipping the Fluorescence Turret Disk
Insert the filter and reflector cubes in the follow-
ing manner:
• Remove the Analyzer (Fig. 23.1).
• Remove the front cover (Fig. 23.2) by pulling it
toward the front.
• Insert the filter or reflector cube into the
mounting in front of you.
To do so, place the filter or reflector cube on
the right side and press it to the left into the
mounting.
Note:
The numbering is located directly below the
holder.
• Apply the included adhesive labels (Fig. 23.3)
corresponding to the installed parts to the
front of the fluorescence illuminator.
• When all filters and reflector cubes have
been inserted, close the front cover plate
again. Ensure that the cover snaps into place.
Fig. 21
Filter cube
front side
Fig. 22
Filter cube
back side
6. Assembling the Microscope
1
2
Fig. 23
Fluorescence Axis
1
Analyzer
2
Front Cover
3
Labels
Fig. 24
Fluorescence Axis without Front Cover
Fig. 25
Mounting the Filter cube
3