IndICaTIOn
The
TANT is intended to shunt cerebrospinal fluid
(cSf) from the lateral ventricles of the brain into
the peritoneum.
13
TECHnICal dESCRIPTIOn
13
14
The
14
TANT is intended for use as a supple-
15
mentary
16
monitoring
16
effect in conjunction with either a conven-
16
tional or adjustable differential pressure
17
valve. Particular attention was paid to de-
18
velop a device that can forestall negative
18
effects such as changes in subcutaneous
18
pressure (e.g. owing to tissue capsules or an
18
unfavorable patient position).
18
18
19
1 titanium casing
19
2 tantalum ball
19
3 sapphire ball
19
4 ball seat
19
5 inlet port
20
6 outlet port
21
Fig. 1: Schematic cross section of the paediSHUNTASSISTANT/SHUNTASSISTANT
paediSHUNTASSISTANT/SHUNTASSIS-
paediSHUNTASSISTANT/SHUNTASSIS-
valve,
in
order
to
allow
for
and
control
of
the
siphon
InstructIons for use |
The paediSHUNTASSISTANT utilises the same
mechanism as the SHUNTASSISTANT. ho-
wever, it is considerably smaller and its func-
tions are intended to meet the special needs of
pediatric hydrocephalus.
TANT consists of a robust titanium casing (1),
a weight-ball (2), as well as a small sealing ball
(3) that rests on the ball seat (4). The cerebro-
spinal fluid flows in through the inlet port (5)
and flows out by way of the outlet port (6).
5
1
2
3
4
6
GB