Technical Data; General Technical Data - Seca 685 Mode D'emploi Et Garantie

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 109

11.TECHNICAL DATA

11.1 General technical data

Dimensions of scale
• Depth
• Width
• Height
Dimensions of weighing platform
• Depth
• Width
• Height
Net weight
Temperature range
• Operation
• Storage
• Transport
Air pressure
• Operation
• Storage
• Transport
104 •
If you want to check whether the scale has been prop-
erly verified, proceed as follows:
1. If necessary, turn off the scale.
2. Press and hold any key and start the scale.
The current content of the verification counter
flashes in the display for a few seconds.
3. Compare the content of the verification counter
output with the number given on the verification
counter sticker.
2
Both numbers have to match for the verification to be
checksum
valid. If the sticker and the verification counter do not
match, the scale must be verified. Please contact your
service partner or seca Service. Once the scale has
been verified, a new, updated verification counter sticker
is used to identify the status of the verification counter.
The person authorized to perform the verification
secures this sticker with an additional seal. The verifica-
tion counter sticker can be ordered from seca Service.
General technical data
seca 685
965 mm
910 mm
1100 mm
850 mm
800 mm
57 mm
59 kg
+10 °C to +40 °C
-10 °C to +65 °C
-10 °C to +65 °C
700-1060 hPa
700-1060 hPa
700-1060 hPa