7. Wipe the support surface covers with a clean, dry cloth to remove any excess liquid or disinfectant.
8. Allow the support surface covers to dry completely before returning to service.
C C o o v v e e r r r r e e p p l l a a c c e e m m e e n n t t , , t t o o p p
T T o o o o l l s s r r e e q q u u i i r r e e d d : :
Zipper pull tool or equivalent
P P r r o o c c e e d d u u r r e e : :
1. Apply the I I n n T T o o u u c c h h brakes.
2. Unplug the support surface power cord.
3. Remove the headboard and footboards and set aside.
4. Lower all the siderails.
5. Unsnap the two secure snaps (C) at the foot end corners (Figure 10).
6. Unsnap the four corner retainers (B) (Figure 10).
7. Insert a zipper pull tool or equivalent through the hole in the zipper slider to unlock.
8. Using the zipper pull tool or equivalent, unzip the support surface top cover from the bottom (A) (Figure 10).
N N o o t t e e - - The zipper starts on the patient right side near the foot end.
9. Remove and discard the top cover.
10. Reverse to install the new cover.
N N o o t t e e
• After installation, remove the zipper pull tool or equivalent from the zipper.
• Connect the corner retainers. Place the strap below each D-ring and snap to the cover. Repeat for the other three
corners.
• Cover the zipper with the support surface cover watershed.
11. Verify proper operation before you return the product to service.
EN
20
A
F F i i g g u u r r e e 1 1 0 0 – – T T o o p p c c o o v v e e r r
B
C
2972-009-001 AB.0