N N o o t t e e - - The reverse side of the support surface is shown for reference. Make sure that the support surface is installed
with the top cover facing up.
6. Put the retainer strap through both D-Rings (C), both sides.
7. Thread the retainer strap through the bottom D-Ring, both sides.
8. Pull the retainer strap until the loop around the Fowler is tight, both sides.
9. Insert the loose end of the retainer straps behind the Fowler.
10. Reinstall the I I n n T T o o u u c c h h headboard and footboard.
11. Position the support surface against the footboard. Allow a small air gap at the headboard.
12. Connect the integration cable from the support surface foot end to the I I n n T T o o u u c c h h cable adapter (B) (Figure 2).
13. Turn the locking collars clockwise to secure the connection.
14. Place the linens on the support surface as loose as possible to avoid creating a h h a a m m m m o o c c k k i i n n g g effect and increasing
interface pressures.
15. Plug the power cord for the support surface into the I I n n T T o o u u c c h h foot end auxiliary power outlet (A) (Figure 2).