• *A zipper pull tool or equivalent is required to access the internal components of the support surface.
• **You cannot perform these preventive maintenance checks without the I I n n T T o o u u c c h h footboard. See the I I n n T T o o u u c c h h
integrated Models 2131 and 2141, software version 4.0 or higher manual.
• The zipper overlaps at the end and may appear to be misaligned.
Replace the following items every two years:
Cover assembly, top and bottom
Product serial number:
Completed by:
Date:
C C l l e e a a n n i i n n g g
W W A A R R N N I I N N G G
• Always unplug the support surface power cord before cleaning or disinfecting to avoid the risk of shock.
• Do not immerse the support surface or foot box in cleaning or disinfectant solutions.
• Do not allow fluid to pool on the support surface. Fluids may cause degradation of components and may cause
unpredictable safety and performance of this product.
• Always inspect the support surface covers (top and bottom) for tears, punctures, excessive wear, and misaligned
zippers as you clean the covers. If the cover is compromised, remove the support surface from service immediately and
replace the cover to prevent cross-contamination.
• Always perform preventive maintenance based on the usage level of the product. An increase in usage which may
include more frequent cleaning and disinfection, may adversely affect the life of the support surface.
• Always make sure that you wipe each product with clean water and thoroughly dry each product after cleaning or
disinfecting. Some cleaning and disinfecting agents are corrosive in nature and may cause damage to the product if you
use them improperly. If you do not properly rinse and dry the product, a corrosive residue may be left on the surface of
the product that could cause premature degradation of critical components. Failure to follow these cleaning or
disinfecting instructions may void your warranty.
C C A A U U T T I I O O N N
• Do not iron, dry-clean, or tumble dry the support surface covers.
• Do not power wash the support surface.
• Always completely dry the support surface covers before storing, adding linens, or placing a patient on the surface.
P P r r e e r r e e q q u u i i s s i i t t e e : : Minimum of two operators are required to clean the bottom support surface cover.
Always follow hospital protocol for cleaning and disinfecting.
To clean the support surface covers between patient uses, follow these steps in order:
1. If powered, unplug the I I n n T T o o u u c c h h from the wall outlet and turn the battery switch to off.
2. Unplug the support surface power cord.
EN
18
F F i i g g u u r r e e 9 9 – – G G r r o o u u n n d d l l u u g g ( ( p p a a t t i i e e n n t t l l e e f f t t s s i i d d e e ) )
2972-009-001 AB.0