The product must reach the ambient operating temperature before installation or operation to avoid the risk of permanent
damage to this product. Upon initial installation, reset the CPR.
W W A A R R N N I I N N G G
• Risk of Shock. This product must only be connected to a supply mains with protective earth. Grounding reliability can be
achieved only when a hospital grade receptacle is used.
• Explosion risk. This product is not suitable for use in the presence of flammable anesthetic mixture with air or with
oxygen or nitrous oxide other than nasal, mask type, or half bed-length tent type.
• Do not move the support surface without assistance.
• Do not operate the support surface without the I I s s o o l l i i b b r r i i u u m m top and bottom covers and linens.
• Always position the support surface on the I I n n T T o o u u c c h h bed frame to that the support surface does not contact the
headboard.
• Do not use when gaps between the support surface and the bed frame are present to prevent the risk of entrapment.
Only use the I I s s o o l l i i b b r r i i u u m m support surface with I I n n T T o o u u c c h h bed frames.
• Do not use fitted sheets with this support surface.
• Do not attach the power cord to any moving parts of the I I n n T T o o u u c c h h bed frame. Improper routing of the power cord could
cause the risk of electric shock, pinching the power cord, or interfere with bed functions.
• Always confirm proper functionality of the scale system after you install the support surface to prevent product
malfunction.
• Do not use extension cords with I I s s o o l l i i b b r r i i u u m m . It is only intended to be powered by I I n n T T o o u u c c h h with the power cord supplied.
• Always use the power source provided by the I I n n T T o o u u c c h h foot end auxiliary power outlet with I I s s o o l l i i b b r r i i u u m m .
N N o o t t e e - - Do not plug into the power source until the assembly is complete.
P P r r e e r r e e q q u u i i s s i i t t e e : Minimum of two operators are required to install the support surface.
1. Put the I I n n T T o o u u c c h h bed into the flat position before installing the support surface.
2. If powered, unplug the I I n n T T o o u u c c h h from the wall outlet and turn the battery switch to off.
3. Remove the I I n n T T o o u u c c h h headboard and footboard from the bed.
4. Position an operator at each end of the support surface and lift onto the I I n n T T o o u u c c h h bed frame.
5. Thread the retainer straps (A), down through the I I n n T T o o u u c c h h Fowler retainer holes (B) (Figure 1).
2972-009-001 AB.0