W W A A R R N N I I N N G G - - Do not service or perform maintenance while the product is in use.
At a minimum, check all items listed during annual preventive maintenance for all Stryker Medical products. You may need
to perform preventive maintenance checks more frequently based on your level of product usage. Service only by qualified
personnel.
Remove product from service before you perform preventive maintenance inspection.
N N o o t t e e - - Clean and disinfect the exterior of the support surface before inspection, if applicable.
Inspect the following items:
Zipper, snaps, and covers (top, bottom, and fire barrier) are free of tears, cuts, holes, or any other damage
N N o o t t e e - - If excessive wear is observed on the cover or the fire barrier, it is strongly recommended to replace the cover.
Support surface cover labels are legible, have proper adherence, and free of damage.
Handles are free of rips or cracks
Power cord, integration cable, and plug are free of damage
**Perform a functional test of the Isolibrium functions
**Perform system diagnostics and confirm there are no errors
*Internal components for signs of staining from fluid ingress or contamination by fully unzipping the cover
Pods are not leaking or cracking
Hose connections to the manifold and sensor tubes (both ends) are seated fully
Foam is free from large tears or large gouges
Turn bladders are free of excessive wear and function properly
Low Air Loss fans rotate freely and are free of debris and dust
Left and Right CPR releases function properly
Both foot box cooling fans rotate freely and are free of debris and dust
Pump spring mount has visible clearance from the foot box bottom
Listen for abnormal pump vibration and noise level
Clean the foot box intake metal screen
Clean the mesh on the bottom cover
Clean the fan filter screen on the foot box
Standoffs (feet of the foot box) are present
Cable tie is present on the power cord
Current leakage not more than 200 μA
Ground impedance not more than 100 mΩ (Figure 9)
N N o o t t e e - - The ground lug can be accessed through the slit in the cover for the CPR, patient left.
2972-009-001 AB.0