O O p p e e r r a a t t i i o o n n
A A p p p p l l y y i i n n g g t t h h e e l l i i n n e e n n s s
W W A A R R N N I I N N G G
• Always make sure that the operator has access to the CPR straps.
• Do not stick needles into a support surface through the support surface cover. Holes may allow body fluids to enter the
inside (inner core) of the support surface and could cause cross-contamination, product damage, or product
malfunction.
• Do not use fitted sheets with this support surface.
To apply the linens:
1. Apply the linens using the "D" rings for the flat sheet.
2. Thread the four linen corners through the "D" rings (C) attached to the bottom cover to secure the linens (A) to the
support surface (B).
N N o o t t e e - - Do not pull the linens tight. Keep the linens loose and as smooth as possible on top of the support surface to
effectively use the Turn Assist and Lateral Rotation functions.
P P o o s s i i t t i i o o n n i i n n g g a a p p a a t t i i e e n n t t o o n n t t h h e e s s u u p p p p o o r r t t s s u u r r f f a a c c e e
W W A A R R N N I I N N G G
• Always center the patient on the support surface. Align the patient's head toward the headboard before starting
functions. Check the patient frequently to make sure that you maintain the proper positioning.
• Always make sure that the tubing and wiring that is connected to the patient is long enough, stable, and secure during
Lateral Rotation or Turn Assist.
• Always raise all of the I I n n T T o o u u c c h h bed siderails before beginning Turn Assist or Lateral Rotation.
• Do not exceed the safe working load of the I I s s o o l l i i b b r r i i u u m m support surface. Excess weight could cause unpredictable safety
and performance of this system.
• Always use extra caution when reading radiology images taken of a patient on a support surface because internal
components can cause artifacts and distort readings.
C C A A U U T T I I O O N N
• Do not allow sharp objects to come into contact with the support surface that could puncture, tear, or cut the cover.
• Do not allow sharp edges from the X-ray plate to come in contact with the support surface cover. The recommendation is
you cover the X-ray plate with a pillow case or other device before placement under the patient. If damaged, remove the
support surface cover from service immediately to prevent cross contamination.
EN
12
B
F F i i g g u u r r e e 3 3 – – A A p p p p l l y y l l i i n n e e n n s s
A
C
2972-009-001 AB.0