Cautions
(Indicate that equipment or other property may be damaged or may malfunction by misuse)
1. Damage (mechanical and electrical) may result if the unit is dropped or struck against another
object.
2. Thoroughly inspect all electrical cables and probes before each use. Do not use if there is any
evidence of deterioration. Replace if any damage or excessive wear is observed.
3. Using the wrench provided, ensure that the probe is tightened on to the transducer before
operation. Before attaching, be sure probe and transducer threads are clean.
4. Do not allow fluids to enter the cabinet. Damage can result.
5. Continuous irrigation and suction must be applied to prevent overheating of the transducer and
probe. Clogs within the probe or transducer must be cleared immediately.
6. Do not allow the probe to press against the scope during the procedure. Pressure on the probe
may stall the transducer and heat the probe within the scope's working channel.
Section 1.0
Unpacking and Initial Inspection
Proper care and maintenance are critical for safe and effective operation of sophisticated medical
equipment. We recommend careful inspection of all equipment upon receipt and prior to each use as a
safeguard against possible injury to patient or operator.
1. To avoid inadvertent damage, study this manual thoroughly before handling assembling, testing,
using or cleaning the CyberWand unit.
2. Examine the shipping carton and instrument for signs of damage. Any breakage or other
apparent damage should be noted, the evidence retained, and the carrier or shipping agency
notified.
3. Verify that the shipping carton contains the items listed below:
a. CW-USLG
Generator
b. CW-USLT
Transducer
c. CW-USLPC
Power cord
d. CW-USLFS
Foot switch
e. CW-USLRBP
Renal/bladder probe sets (2) (non-sterile)*
* CW-RBPBX
3-pack, Sterile Renal/Bladder Probes sold separately
f. CW-USLW
Wrench
g. CW-USLCS
Cleaning Stylet
h. 80834
Instructions for Use
4. Notify Gyrus ACMI Customer Service immediately if any damage or discrepancies are noted.
CyberWand Dual Ultrasonic Lithotripsy System • Instructions for Use