Front Panel; Back Panel; Footswitch; Ref Cw-Uslfs) - Gyrus Acmi CyberWand CW-USLS Mode D'emploi

Système de lithotritie ultrasonique
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 18
1
5
2
3
4
1
2
1
3
8
General Description of CyberWand Unit and accessories
6
2
3
2.5

Footswitch (

REF
1. LARGE Pedal Mode (gray)
Depressing LARGE pedal causes the CyberWand system to produce maximum pulse energy to
fragment large stones into smaller stone pieces.
2. SMALL Pedal mode (green)
Depressing SMALL pedal causes the CyberWand system to produce a pulse energy level that
keeps the probe in contact with the smaller stone fragments allowing greater fragmentation.
Small stones are repulsed less when the SMALL pedal is used.
4
3. Connector plug
4. Cable
2.3

Front Panel

1. Power Indicator
Power LED is illuminated when power is applied to the system.
O
I indicates power is "ON,"
2. Tune Indicator
Tune LED, flashing when power is applied, indicates that the
CyberWand system is not tuned and is not ready for use.
3. Ready Indicator
When Ready LED is steady on, the system is tuned and ready for
use.
4. Fault Indicator
Fault LED is illuminated when either the footswitch or the
transducer is disconnected.
5. Footswitch Receptacle
Align the white dots to connect the footswitch to the generator
6. Transducer Receptacle
Align the red dots to connect the transducer to the generator.
2.4

Back Panel

1. Cooling Fans (2) with finger guard
2. Power Receptacle
UL Listed 3-prong hospital grade plug
3. Fuse Holder
CAUTION: Do not block fan during operation; unit will overheat.
CW-USLFS)
indicates power is "OFF."