Section 6.0
Operating and storage environment
Operating environment Ambient temperature
Relative humidity
Air Pressure
Storage environment
Ambient temperature
Relative humidity
Air Pressure
Section 7.0
Preventive Maintenance
7.1

General Checks

The operator should ensure the generator and all cables and all components on transducer are
undamaged prior to beginning any procedure.
7.2

Periodic Safety Checks

The safety checks should be performed at least every 24 months. The inspection can only be
performed by a person who has adequate training. knowledge, and practical experience. All the
equipment used in safety check must be calibrated. Gyrus ACMI recommends returning the unit for
safety inspection or contact your local representative for information about maintenance contracts.
Section 8.0

Troubleshooting

8.1
Troubleshooting
Irregular Description
Possible cause
Power indicator LED is not
The power cord is not securely
illuminated.
plugged into a receptacle
The power cord is not securely
plugged into generator
Power switch is not turned ON
The fuse has blown
Fault Indicator light is
Footswitch is not plugged into
illuminated.
generator
Transducer is not plugged into
generator
Poor performance or low
• Mis-tuned probe
power
• Loose Probe
8.2

Fuse Replacement

Warning:
Ensure that the power cord is disconnected from the instrument before opening the fuse holder.
1. Remove the fuse holder with a flat screwdriver, to expose fuse
2. Replace the blown fuse with a new fuse (TYPE T 250V 5A).
3. Replace the fuse holder securely into the generator.
10 – 40°C
30% – 70%
70 – 106 kPa
-25 – +55°C
10% – 85%
70 – 106 kPa
Solution
Connect power cord into receptacle.
Note: Ensure the power switch is at OFF
position before plugging in.
Connect power cord into generator.
Note: Ensure the power switch is at OFF
position before plugging in.
Turn the switch ON.
Replace fuse as described in 8.2 Fuse
Replacement below.
Plug footswitch into generator.
Note: Ensure the power switch is at OFF
position before plugging in.
Plug transducer into generator.
Note: Ensure the power switch is at OFF
position before plugging in.
• Re-tune system (Section 3.1).
• Retighten probe and re-tune system.
80834 REV ED