12. Technical data
12.1 Classification acc. to Directive 93/42/EEC
Art. no.
Designation
GA340
Dermatome
GA341
0.1 mm Dermatome
12.2 Performance data, information about standards
Max. power
Max. oscillation frequency
Blade travel
Cutting depth setting
Cutting width
Weight (operational)
Dimensions (L x W x H, ready for
operation)
Applied part
EMC
Conforming to standard
The product was subject to 500 preparation cycles as a test by the manu-
facturer and has passed.
12.3 Operating mode
Operating mode
approx. 250 W
-1
6 500 min
approx. 3.1 mm
0.2 mm to 1.2 mm (with GA340)
0.1 mm to 1.1 mm (with GA341)
Scale gradations of 1/10 mm
max. approx. 78 mm, set by
10 flaps of width 8 mm
1.65 kg ±10 %
277 mm x 113 mm x 83 mm ±5 %
Type BF
IEC/DIN EN 60601-1-2
IEC/DIN EN 60601-1
Operation with non-periodic load and
speed changes (type S9 pursuant to
IEC EN 60034-1)
60 second application, 60 second
pause
10 repetitions
30 min cooling time
Max. temperature 48 °C
12.4 Environmental conditions
Temperature
Relative
Class
humidity
IIa
Atmospheric
IIa
pressure
13. Disposal
Note
The user institution is obliged to process the product before its disposal, see
Validated reprocessing procedure.
Adhere to national regulations when disposing of or recy-
cling the product, its components and its packaging!
The recycling pass can be downloaded from the Extranet
as a PDF document under the respective article number.
(The recycling pass includes disassembling instructions for
the product, as well as information for proper disposal of
components harmful to the environment.)
Products carrying this symbol are subject to separate col-
lection of electrical and electronic devices. Within the
European Union, disposal is taken care of by the manufac-
turer as a free-of-charge service.
Detailed information concerning the disposal of the product is avail-
able through your national B. Braun/Aesculap agency, see Technical
Service.
14. Distributor in the US/Contact in Can-
ada for product information and com-
plaints
Aesculap Inc.
3773 Corporate Parkway
Center Valley, PA, 18034,
USA
Operation
Storage and transport
10 °C to 27 °C
-10 °C to 50 °C
30 % to 75 %
10 % to 90 %
700 hPa to 1 060 hPa
500 hPa to 1 060 hPa