Aesculap
Dermatome GA340 / Dermatome 0.1 mm GA341
7.9
Automatic cleaning/disinfection with manual pre-cleaning
Note
The cleaning and disinfection device must fundamentally have a tested
efficacy (such as FDA approval or CE label pursuant to DIN EN ISO 15883).
7.9.1
Manual pre-cleaning with a brush
Phase
Step
I
Rinsing
II
Brushes
D–W:
Drinking water
RT:
Room temperature
►
Do not clean the product in a ultrasonic bath and do not immerse the
product in any fluids. Let any fluid incursions drain out immediately,
otherwise there is a danger of corrosion and loss of function.
Phase I
►
Dismantle the flap rod for cleaning. Store flap rods, flaps and nut sep-
arately in tray.
►
Mobilize non-rigid components during cleaning.
►
Thoroughly clean the product under running water.
7.9.2
Mechanical alkaline cleaning and thermal disinfection
Machine type: single-chamber cleaning/disinfection device without ultrasound
Phase
Step
I
Pre-rinse
II
Cleaning
III
Intermediate rinse
IV
Thermal disinfection
V
Drying
D–W:
Drinking water
FD–W:
Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least)
*Recommended: BBraun Helimatic Cleaner alkaline
12
Acculan 4
®
T
t
[°C/°F]
[min]
RT (cold)
-
RT (cold)
-
T
t
[°C/°F]
[min]
<25/77
3
55/131
10
>10/50
1
90/194
5
-
-
Note
The cleaning and disinfection machine used for processing must be ser-
viced and checked at regular intervals.
Conc.
Water
Chemistry/Note
[%]
quality
-
D–W
Until visually clean
-
D–W
Until visually clean
Phase II
►
Mobilize non-rigid components during cleaning.
►
Brush difficult to access surfaces with a suitable plastic cleaning brush
for at least 1 min.
►
After manual preparation, check visible surfaces for residue and repeat
the pre-cleaning process as needed.
Note
For details on difficult to access surfaces, see Acculan pre-cleaning and
care information TA016000 (available in Aesculap Extranet at
https://extranet.bbraun.com).
Water
Chemistry/Note
quality
D–W
-
■
FD–W
Concentrate, alkaline:
– pH ~ 13
– <5 % anionic surfactant
■
0.5 % working solution
– pH ~ 11*
FD–W
-
FD–W
-
-
mind. 10 min at max. 120 °C