Aesculap
Dermatome GA340 / Dermatome 0.1 mm GA341
8.
Maintenance
To ensure reliable operation, the product must be maintained in accor-
dance with the maintenance labeling or at least once a year.
YYYY-MM
For technical service, please contact your national B. Braun/Aesculap
agency, see Technical Service.
9.
Troubleshooting list
►
Have defective products repaired by Aesculap Technical Service, see Technical Service.
Issue
Product not running
Product becomes too hot Excessive use
Insufficient power
Loud running noise
14
Acculan 4
®
Cause
No battery
Battery not charged
Battery defective
Product in locked OFF position Safety catch is in OFF position
Product defective
Reprocessing/maintenance
carried out incorrectly
Fall damage, product defec-
tive
Blunt tool
Product defective
Blunt tool
Defective product drive/ball
bearings
Detection
No battery in the battery shaft
No signal tone when inserting the bat-
tery
No signal tone when inserting the bat-
tery
Product not running
Product heating
Product heating
Product heating
Heating of tool and product
Insufficient product performance
Large amount of heat generated after a
short time
Tool blades worn out
Loud, perceptible noise during opera-
tion
Remedy
Insert battery.
Charge battery in charger.
Have the manufacturer repair the battery.
Turn the safety catch into the ON position.
Have the manufacturer repair the product.
Follow operating instructions (operating
mode).
Follow operating instructions (preparation,
care).
Preventative: lubricate product before every
sterilization.
Have the manufacturer repair the product.
Change tool.
Follow operating instructions (preparation,
care).
Preventative: lubricate product before every
sterilization.
Have the manufacturer repair the product.
Follow operating instructions (operating
mode).
Have the manufacturer repair the product
Change tool.
Follow operating instructions (preparation,
care).
Preventative: lubricate product before every
sterilization.
Have the manufacturer repair the product