Aesculap
Dermatome GA340 / Dermatome 0.1 mm GA341
7.3
Preparations at the place of use
Remove all attached components from the product (dermatome blade,
battery, accessories).
Remove any visible surgical residues as much as possible with a damp,
lint-free cloth.
Place the dry product in a sealed waste container and forward it on for
cleaning and disinfection within 6 hours.
7.4
Disassembling the product before carrying out the
reprocessing procedure
Remove battery 7 from the product 1, see Removing the rechargeable
battery.
Open the clamping levers 15 to remove the dermatome blade 11, see
Removing the dermatome blade.
Disassemble the product 1 immediately after use, according to instruc-
tions.
7.4.1
Removing the flap rod of the dermatome
Loosen nut 26 by turning it clockwise (left-handed thread).
Screw the nut 26 back to the end of the visible surface.
Press on nut 26 and push flap rod 23 approx. 4 mm to one side.
Twist flap rod 23 until it can be removed.
Remove flap rod 23.
Slide off the flaps 24 from flap rod 23.
The flap rod 23 has been dismantled.
7.5
Preparation before cleaning
Prior to the first mechanical cleaning/disinfection process, mount the
ECCOS fixations in a suitable tray (e.g. JC254R) or use the ECCOS tray
GB256R with fixations equipped.
Insert the products in the correct position into the ECCOS fixations, see
Fig. G.
7.6
Single-use products
Dermatome blade GB228R
Risk of infection for patients and/or users and
impairment of product functionality due to reuse.
Risk of injury, illness or death due to contamination
WARNING
and/or impaired functionality of the product!
Do not reprocess the dermatome blade GB228R.
10
Acculan 4
®
7.7
Cleaning/disinfection
7.7.1
Product-specific safety instructions for the reprocessing pro-
cedure
Damage to the product due to inappropriate clean-
ing/disinfecting agents and/or excessive tempera-
tures!
CAUTION
Use cleaning and disinfecting agents according
to the manufacturer's instructions which
– are approved for plastic material and high-
grade steel,
– do not attack softeners (e.g. in silicone).
Do not use cleaning agents that contain ace-
tone.
Observe specifications regarding concentration,
temperature and exposure time.
Do not exceed the maximum temperature of
60 °C during chemical cleaning and/or disinfec-
tion.
Do not exceed the maximum temperature of
96 °C during thermal disinfection with FD
water.
Dry the product for at least 10 minutes at a
maximum of 120 °C.
Damage or destruction of the batteries due to pro-
cessing!
Protect battery from moisture.
CAUTION
Note
The indicated drying time is a guide time only. It must be checked taking
into account the specific conditions (e.g. load) and if applicable adjusted.