Issue
Cause
Lid cannot be
Lid not compatible
attached/removed
Lid deformed/defective
Connection to product
deformed/defective
Battery cannot be
Battery not compatible
inserted/removed
Battery deformed/defective
Battery shaft to product
deformed/defective
Sterile funnel or battery
Sterile funnel or battery
removal device not
removal device not compati-
attachable
ble
Sterile funnel or battery
removal device
deformed/defective
Battery shaft to product
deformed/defective
Speed control knob not
Product in locked OFF position Safety catch is in OFF position
working
Speed control knob sticks/is
defective
Dermatome blade not
Defective gearings
moving
Insufficient cutting per-
Dermatome blade blunted
formance of dermatome
Adjusted dermatome
blade
Cutting depth
Skin grafts uneven / too
uneven/deviating from
thick / too thin
setting
Detection
Lid does not click into place
Lid is difficult or cannot be
mounted/removed
Lid is difficult or cannot be
mounted/removed
Battery cannot be completely inserted
into the battery shaft
Battery is difficult or cannot be
mounted/removed
Battery is difficult or cannot be
mounted/removed
Sterile funnel or battery removal device
not attachable to the battery shaft
Sterile funnel or battery removal device
difficult or not attachable
Sterile funnel or battery removal device
difficult or not attachable
Speed control knob not working
Product runs loudly
Worn cutting edge of dermatome blade Replace dermatome blade
Product has insufficient power
Mechanical influence (e.g. fall on floor) Have the manufacturer repair the product
Remedy
Use the compatible lid for GA340/GA341.
Have the manufacturer repair the lid.
Have the manufacturer repair the product.
Use the compatible battery for GA340/GA341.
Have the manufacturer repair the battery.
Have the manufacturer repair the product
Use compatible sterile funnel or battery
removal device for GA340/GA341.
Change sterile funnel or battery device.
Have the manufacturer repair the product.
Turn the safety catch into the ON position.
Have the manufacturer repair the product.
Have the manufacturer repair the product
Have the manufacturer repair the product
15