PH value (degree of acidity)
Appearance
Hardness (E Ions of alkalin earth)
PAR. 10
In case the pump does not work, the loading tank can be filled by the operator in the following way:
1. Turn on the device.
2. Unscrew the cap.
3. Insert the supplied pipe for manual filling outside of which you can find a funnel.
4. Pour the distilled water in the funnel which is kept higher thank the loading cap;
5. Pour up to when the pilot light "LEVEL WATER TANK, max." is turned on.
6. Screw the cap checking its tightening.
PAR. 11
The device is fitted with a tank for collecting the waste waters used during the sterilisation cycle; this tank is
separated from the loading tank.
When the discharge tank is full, the max level pilot light "EXHAUST TANK", is turned on.
To discharge the tank, proceed as follows:
•
Connect the supplied pipe to the discharge tap.
•
Unscrew the discharge tap in counterclockwise direction.
•
Flow the waste waters.
•
Screw the discharge tap without forcing it and checking it is closed properly.
•
In case of sterilization failure, the used water in discharge tank, can have contamined residual. This
water must de disposal in the black water, use protettive latex glover to carry out the operation.
The device does not work in case the discharge tank is full. In case the operator tries to start a cycle even in
this state, the device locks visualising "ER3" on the "TIME/MESS". display.
PAR. 12
COMMISSIONING AND PREPARING THE STERILIZATION
1. Carry out the installation operations as described in PAR.6, the autoclave is ready to be used. Pay
attention to the instructions below.
2. Turn on the autoclave pressing the main switch. The autoclave pre-heats at 80°C; if it is not used, it
remains in the stand-by cycle for 60 minutes.
3. Open the door using the lever. In case you cannot open the door, check that the door lock in not
inserted. Pressing the "14A" to operate the door interlock, to carry out the door release.
4. Load the instruments placing them on the supplied trays. It is fundamental not to overlap the instruments
or the bags and not to overload the trays. The max. capacity is of 700 g for each tray.
5. To obtain a good drying it is important that the wrapped instruments are placed in the trays with plastic
side downwards and obviously the paper side upwards. This is done to ease the output of air bubbles
during sterilisation and the vapour drops condensed during drying, as specified in PAR.7.
6. Each time a sterilisation cycle is carried out insert an integrator in the sterilisation chamber to have the confirmation
of the carried out sterilisation; we recommend to place the integrator in the middle of the sterilisation chamber. It is
MANUAL FILLING OF THE WATER TANK
EMPTYING THE TANK
Colourless clean without sediment
22
5 to 7
0.02
mmol/l
DNVT008 Rev_5