Sterilization
After performing the cleaning instructions specified above, perform one of the
following sterilization cycles. Note: for necessary materials and equipment, see the
Materials and Equipment table.
Steam
1.
Clean and prepare the device as recommended in this user manual.
2.
If using a sterilization tray, follow any additional instructions provided with
the tray. Use only trays that are compatible with steam sterilization.
3.
Double wrap the device (or tray) prior to sterilization.
4.
Sterilize the device using the parameters below.
Note: The water used in the autoclave process must meet standards for
clean steam per AAMI ST79 Annex M – Steam Quality.
5.
After sterilization, allow the cable to completely dry and cool to room
temperature before reuse.
• For routine reprocessing with steam sterilization, use the following
Gravity or Pre-Vacuum cycles.
Gravity
Wrapping
Temperature
Sterilization Time
Dry Time
Pre-Vacuum
Wrapping
Temperature
Sterilization Time
Dry Time
• Immediate-use steam sterilization (IUSS, or "Flash") is intended only
for emergency situations. In the event that immediate-use steam
sterilization is required, the following instructions should be used.
"Flash" Pre-Vacuum
Wrapping
Temperature
Sterilization Time
Dry Time
Immediate-use steam sterilization should only be used in
carefully selected clinical situations (e.g., an instrument needed
for a case falls on the floor and no replacement instrument
is available). IUSS should not be used due to inadequate
inventory.
Double
132 °C (270 °F)
15 minutes
30 minutes
Double
132 °C (270 °F)
4 minutes
30 minutes
132 °C (270 °F)
4 minutes
134 °C (274 °F)
10 minutes
134 °C (274 °F)
3 minutes
None
134 °C (274 °F)
3 minutes
None
EN-11