3.
Rinse
• Remove the device from the prepared detergent. Rinse the device with reverse
osmosis/de-ionized (RO/DI) water at ambient temperature for 90 seconds or
until all visible detergent residue is removed.
• Flush any mated surfaces and crevices 5 times. After all visible detergent
residue is removed, continue to rinse for 30 seconds.
• Drain excess water from the device and dry it with a clean cloth or filtered
pressurized air.
• Visually inspect the device for cleanliness, paying close attention to hard-to-
reach areas. If visible soil remains, repeat steps 1 through 3.
4.
Soak
• Prepare a non-enzymatic detergent according to the manufacturer
instructions.
• Fully immerse the device in the detergent. Use a syringe to inject 50 mL of the
detergent into any mated surfaces and crevices.
• Soak the device in the detergent for 1 minute.
5.
Brush
• Thoroughly brush the exterior of the device with a soft-bristle brush for
90 seconds, focusing on any mated surfaces or crevices.
• Use a syringe to inject 50 mL of the detergent into any mated surfaces and
crevices 5 times.
6.
Rinse
• Remove the device from the prepared detergent. Rinse the device with reverse
osmosis/de-ionized (RO/DI) water at ambient temperature for 90 seconds or
until all visible detergent residue is removed.
• Flush any mated surfaces and crevices 5 times. After all visible detergent
residue is removed, continue to rinse for 30 seconds.
• Drain the excess water from the device and dry it with a clean cloth or filtered
pressurized air.
• Visually inspect the device for cleanliness, paying close attention to hard-to-
reach areas. If visible soil remains, repeat steps 1 through 6.
EN-8