Materials and Equipment
All materials and equipment required to reprocess the cable shall be supplied by the
user unless otherwise noted.
Item
All phases
Gloves, eye protection, etc.
Cleaning
Water basin
Enzymatic detergent
Tap water
Syringe
2
Soft-bristle brush
Reverse osmosis/deionized
water
4
Clean cloth or filtered
pressurized air (≤40 psi)
Non-enzymatic detergent
Automated washer
Disinfection
Water basin
Disinfecting solution
Tap water
Soft sponge
Deionized water
7
Sonicator
Clean cloth or filtered
pressurized air (≤40 psi)
Washer disinfector
Sterilization
Sterilization system
Sterilization wrap
Sterilization tray
9
EN-6
Description
Wear protective equipment as required by the
medical facility and procedure.
Large enough to accommodate the device
Used in cleaning solution to remove surgical debris
1
To prepare cleaning solutions
To inject detergent into hard-to-reach areas of the
device
To clean exterior of the device
3
To rinse the device
To assist with drying
Used in cleaning solution to remove surgical debris
5
For using the automated cleaning procedure
Large enough to accommodate the device
≥ 0.55% ortho-phthalaldehyde
6
To prepare disinfecting solution
To wipe the device during manual disinfection
To rinse the device during manual disinfection
To sonicate the device during manual disinfection
To assist with drying
To perform thermal disinfection
• Steam (Autoclave)
• Steris/Amsco® V-PRO® 1, V-PRO 1 Plus, or V-PRO
maX
• Sterrad 100S, 200, NX®, or 100NX
To maintain sterile barrier
8,9
Optional. Must be compatible with sterilization
method.