Stryker AIM SafeLight Mode D'emploi page 13

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 19
Automated Cleaning
Note: for necessary materials and equipment, see the Materials and Equipment
table.
1.
Brush
• Prepare an enzymatic detergent according to the manufacturer instructions.
• Fully immerse the device in the detergent. Thoroughly brush the exterior of
the device with a soft-bristle brush for 90 seconds, focusing on any mated
surfaces and crevices.
2.
Rinse
• Rinse the device with reverse osmosis/de-ionized (RO/DI) water at ambient
temperature for 90 seconds or until all visible detergent residue is removed.
• After all visible detergent residue is removed, continue to rinse for 30 seconds.
3.
Automated wash
• Place the device in the washer on an incline to facilitate drainage.
• Program the washer using the following parameters:
Phase
Pre Wash
Enzyme Wash
Wash 1
Rinse 1
Dry Phase
• Filtered pressurized air can be used to aid in drying.
• Visually inspect the device for cleanliness, paying close attention to hard-to-
reach areas. If visible soil remains, repeat steps 1 through 3.
Recirculation
Time
2 minutes
2 minutes
2 minutes
2 minutes
7 minutes
Temperature
Cold water
Hot water
Set point 66 °C
(151 °F)
Hot water
115 °C (239 °F)
Detergent Type
N/A
Enzymatic Detergent
Non-Enzymatic
Detergent
N/A
N/A
EN-9