Storage
The monopolar electrode handle (incl. connecting cables) should be stored in a cool and dry
place at temperatures between -20°C and +50°C (-4 to 122°F), a relative humidity of 10% to
100% and an atmospheric pressure of 500 to 1060 hPa. Direct sunlight exposure over an ex-
tended period of time should be prevented.
General Information
Please carefully read this user information prior to using the handle.
The instructions given herein are not a substitute for the information provided in the user
manuals for other HF devices and accessories.
Improper handling and cleaning, as well as non-intended use, can lead to premature wear or
life-threatening risks to patients and users.
Warning: Instruments for HF electrosurgery may only be used by persons specially
trained or instructed in their use.
Product Description and Intended Use
The monopolar HF electrode handle is designed for use across all electrosurgical fields. It is
intended for monopolar cutting and coagulating of biological tissue using any type of electrode
with a shaft Ø of 4 mm (± 0.1 mm). An electrode suitable for the specific surgical task is to be
inserted into the tip of the handle. The electrode can be installed and locked in place in differ-
ent positions. It must be pushed in as far as it goes (the hexagonal attachment of the elec-
trode must fully disappear in the handle tip). A suitable cable is provided to connect the HF
electrode handle to the output of the HF generator. The cutting or coagulation current is set on
the HF generator. The current is activated for use via the pushbuttons provided on the handle
(yellow button = activation of cutting current; blue button = activation of coagulation current;
or gray button (when using a handle having only one button) = activation of the mode set on
the HF unit).
In principle, electric energy can be supplied to the handle electrode by footswitch activation as
well, so the handle can also be used in conjunction with the footswitch.
Safety Instructions
Prior to each use, check the handle for damage and verify its proper functioning. Never use
•
defective HF electrode handles!
The handles are made of composite materials and cannot be repaired.
•
Protect the HF electrode handle against mechanical damage. Never throw or drop it! Do not
•
use force!
Hot-air sterilization is not permitted!
•
Use only clean and sterilized HF electrode handles.
•
Do not connect the handle to other types of current source.
•
Run the connecting cable of the handle in such a way that it touches neither the patient nor
•
other cables in the surgical field.
When using HF surgery in conjunction with physiological monitoring units on the patient,
•
the monitoring electrodes should be placed as far away from the HF electrodes as possible.
Do not use needle electrodes for monitoring! Non-observance can result in burns or
electric shock!
6
Instructions for Use
Monopolar Electrode Handle
V 1.0