Do not clean the handle together with sharp-edged or pointed objects.
-
-
The handle can be cleaned in an ultrasonic bath.
Following cleaning/washing, all surfaces and bore holes must be thoroughly rinsed with
-
sterilized water.
Water quality must comply with DIN EN 285, Appendix B.
-
Do not use compressed air > 2 bar for handle cleaning – danger of pushbutton damage.
The drying cycle should not exceed a temperature of +95°C (203°F). Keep the handles
away from heat-generating parts of the dryer, as plastic components of the handle could
melt, thus posing a significant risk of functional damage as well as a serious safety hazard!
Following cleaning, inspect the handle for cleanness. If necessary, repeat the cleaning cycle
or clean manually.
Cleaners and Disinfectants for Use on KLS Martin HF Surgical Accesso-
ries
Be sure to follow the instructions provided by the manufacturer(s) of the cleaning and disin-
fecting agents used. When using machine cleaning, the instructions given by the manufacturer
of your washer-disinfector must be duly observed as well. Please note that Gebrüder Martin
will accept no liability for malfunction in its products caused by disinfectants and the pro-
cessing methods used, nor for the effectiveness of the disinfectants used.
We recommend using cleaners and disinfectants offered by the following suppliers:
Bode Chemie Hamburg
Melanchthonstraße 27
D-22525 Hamburg
Internet:
www.bode-chemie.de
Should you be in doubt about the suitability of your processing chemicals, do not hesitate to
send us a sample – we will be glad to test them for you in our in-house lab.
Sterilization
Steam sterilization in an autoclave (pre-vacuum cycle, validated procedure acc. to
EN ISO 17665) at 134°C (273.2°F) with a holding time of at least 5 minutes.
Sterilize only cleaned products
Use suitable sterilization containers
Hot air or so-called "flash" autoclaving are not permitted for handle sterilization!
To prevent handle damage, the autoclave temperature must not exceed 137°C (278.6°F).
Do not exceed the maximum load permitted for the sterilizer!
It is the operator's responsibility to guarantee the continued sterile condition of the product
after completion of the sterilization process. To this end, a standardized packaging means
must be used, large enough to accommodate the HF electrode handle without creating seal
tension.