Warmig up time and vacuum fractionation can vary, because of autoclave's conditions, between 25 and 30
minutes. You have to an this time to sterilization's times showed in table
9.4 – TEST CYCLES
-Helix test: this test serves to verify steam penetration capacity in a hollow body. Use the Helix Indicator
System.
- Bowie and dick test: this serves to verify steam penetration capacity in a porous load. Use Bowie and Dick
test pack (top table).
- Vacuum test: this serves to verify the preservation capacity of the autoclave vacuum; test report is prepared
by the autoclave printer.
It is preferable to use tests distributed by TECNO GAZ spa.
TEST CYCLES TABLE
CYCLES
134°C Helix/
Bowie&Dick
134°C Vacuum
CYCLE DIAGRAM
STERILIZATION
DRYING TIME
TIME
(Minutes)
( Minutes)
3.5
4
MAX. LOADING FOR
MINIMUM WORK PRESSURE
(Kg)
TRAY
57
( bar, differential pressure )
2.02
-0.9
DXBA408 Rev_14