Cleaning; If You Encounter Problems; Replacing The Seal - Burkert 8011 Manuel D'utilisation

Masquer les pouces Voir aussi pour 8011:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Maintenanceandtroubleshooting
Warning
risk of injury due to non-conforming maintenance.
• Maintenance must only be carried out by qualified and skilled
staff with the appropriate tools.
• Ensure that the restart of the installation is controlled after any
interventions.
9.2.

cleaning

Depending on the nature of the fluid, regularly check for clogging of
the paddle-wheel.
note
the device may be damaged by the cleaning liquid.
• Clean the device with a cloth slightly dampened with water or a
cleaning liquid compatible with the materials the device is made of.
9.3.

if you encounter problems

problem
recommended action
The device does not
• Check the wiring
function
• Check that the device is
energized
The pulse output
Check whether the wiring is
does not work
suitable for the output type,
NPN or PNP
The flow rate meas-
Recalculate and change the
urement is incorrect
setting of the K factor
Type 8011
see chap.
7.4
7.4
6.3.5
9.4.

replacing the seal

O-ring for plastic fitting
Fig. 14: Exploded view of the 8011
Unscrew the 4 screws in the electronic module and remove it
from the fitting.
Remove the used seal.
Clean the surfaces on which the seal is placed.
Insert the new seal (see Fig. 14).
Position the electronic module on the fitting so that the arrow
points in the fluid direction on versions with optical sensor.
Insert the 4 screws into the electronic module (use the long
screws for a plastic S012, DN6 or DN8 fitting).
Tighten the 4 screws in an alternating pattern, to a torque of
1.5 Nm.
English
O-ring for metal fitting
31