Time Weighted Average (TWA)

The TWA flag appears in the upper portion of the display to show
the average exposure since the TWA reading was reset.
When the amount of gas detected by the Orion plus Multigas
Detector is greater than the eight-hour TWA limit, the:
Alarm Sounds and Alarm Lights Flash
TWA flag flashes.
The TWA alarm is calculated over an eight-hour exposure. Calculation
examples are as follows:
1-hour exposure of 50 PPM:
(1 hour x 50 PPM) + (7 hours x 0 PPM)
12-hour exposure of 100 PPM:
(12 hours x 100 PPM)
8 hours
NOTE: The accumulated reading is always divided by eight hours.
If the STEL or TWA alarm condition is reached while using
the instrument as a personal or area monitor, leave the con-
taminated area immediately; the ambient gas concentration
has reached the preset STEL or TWA alarm level. Failure to
follow this warning will cause over-exposure to toxic gases,
which can result in serious personal injury or death.

Turning OFF the Orion plus Multigas Detector

Push and Hold the ON-OFF button for five seconds.
Gas readings end and the Hourglass displays.
NOTE: Releasing the ON-OFF/PAGE button before the five seconds
elapse returns the instrument to the Measure page.
3-12
8 hours
= 150 PPM

" WARNING

= 6.25 PPM