B Status/Error during NIBP Monitoring
"Cuff error"
—cuff is not wrapped correctly, or is not connected
"Air leak"
—Air moving part, tube or the cuff leak air.
"Pressure error"
—Unstable cuff pressure or tangled cuff tubing
"Signal weak"
—Very weak signal because of the cuff, or the patient has very weak pulse
"Over extent"
—The measurement range exceeds 255 mmHg (Pediatric patient over 135 mmHg)
"Over motion"
—The repeated measurement due to moving, excessive noise during the stepping inflation and
"Signal overflow"
—Blood pressure amplifier overflow due to excessive movement
"Leak in gas run"
—Leaking during the pneumatic device testing
"System error"
—Abnormal condition of CPU, such as register overflow, divided by zero
"Adult"
—The blood pressure measuring now is in adult mode. In this case, it is not allowed to monitoring
"Pediatric"
—The blood pressure module is now worked in Pediatric measuring mode.
"PROBE OFF"
—SpO
"LEADS OFF"
—The ECG electrodes or cable fell off
"LEARNING"
—Learning arrhythmia for 15 seconds
"DEMO"
—The monitor is displaying the demo waveforms, which are generated by the monitor itself.
Operator's Manual for Patient Monitor
measuring pressure and pulse, e.g. during patient shaking motion
Pediatric or neonatal patient. Otherwise, there may be serious danger to the Pediatric
monitored.
probe fell off