The operating environment of the equipment must meet the requirements specified in this manual. Otherwise,
unexpected consequences, e.g. damage to the equipment, could result.
The environment where the equipment is used shall be reasonably free from noises, vibration, dust, corrosive,
flammable and explosive substances. If the equipment is installed in a cabinet, sufficient space in front and behind
shall be left for convenient operation, maintenance and repair. Moreover, to maintain good ventilation, the equipment
shall be at least 2 inches (5cm) away from around the cabinet.
When the equipment is moved from one place to another, condensation may occur as a result of temperature or
humidity difference. In this case, never start the system before the condensation disappears.
3.2 Getting Started
3.2.1 Connecting to Power Source
1. Using AC Power Source
Make sure that the AC power supply is (100-240)VAC, 50Hz/60Hz.
Use the power cable provided by the manufacturer. Insert one end of it to the AC power input of the
monitor and the other end to the three-pin outlet of the power source with protected-earth.
To eliminate potential differences, the monitor has a separate connection to the equipotential
grounding system. Connect one end of the provided ground wire to equipotential grounding terminal
on the rear of the monitor, and connect the other end to one point of the equipotential grounding
system.
Caution: 1. Ensure that the monitor is grounded correctly.
2. If you have any doubt to the grounding layout and its performance, you must use the built-in battery to
power the monitor.
After the mains power supply has been interrupted while power switch remains in the "on" status and is
restored after a period of time that is longer than 30 seconds, the monitor will run by the last settings when
restarting the monitor.
The monitor is applicable to connect to the public mains power network.
2. Using Battery
The following steps should be followed to install the battery:
Step 1: open the battery cover;
Step 2: pull out the battery cable and connect it to the battery pack;
Step 3: push the battery pack into the battery compartment and lock it;
Step 4: close the battery cover.
Caution: it's better to recharge the battery after it is used up, and the charging time should be 12~15 hours
long.
Battery life: Provided that a battery is new and fully charged, the minimal working time of the monitor with
accessories connected is declared in the table below:
Operator's Manual for Patient Monitor