17.2 RESP
1. RESP rate measuring range: 0rpm~120rpm
2. RESP rate accuracy: ±5% or ±2 rpm, whichever is greater
3. RESP rate alarm limit setting range: High: 1rpm~150rpm; Low: 0rpm~149rpm.
4. Alarm tolerance: ±1rpm
17.3 TEMP
1. TEMP measuring range: 21.0℃~50.0 ℃
2. TEMP measuring accuracy: ±0.2 ℃ for the range from 25℃ to 45℃, 0.4 ℃ for other ranges.
3. TEMP responding time: ≤150s
4. TEMP minimum measuring time: ≥130s
5. Measuring site: body surface
6. Mode of operation: direct mode
7. Unit: ℃ and °F
8. TEMP alarm limit setting range: High: 0℃~60℃; Low: 0℃~59.9℃.
9. Tolerance: ±0.1℃
17.4 NIBP
1. Measuring method: Oscillometric Technique
2. Pneumatic pressure measuring range: 0 mmHg~300mmHg
3. Accuracy of pressure measurement: ±3 mmHg
4. Typical measurement time: <30 seconds (adult cuff)
5. Measurement time on the average: < 90 seconds
6. Air release time while the measurement is canceled: <2 seconds (typical adult cuff)
7. Initial cuff inflation pressure
Adult: <150 mmHg; Infant: <120 mmHg;
8. Overpressure protection limit
Adult: 300 mmHg; Infant: 240mmHg; Neonate: 150 mmHg
9. NIBP measurement range:
Pressure(unit)
mmHg
SYS
kPa
mmHg
MAP
kPa
mmHg
DIA
kPa
Operator's Manual for Patient Monitor
Neonate: <70 mmHg Tolerance: ±5 mmHg
Adult
40~275
5.3~36.7
5.3~26.7
20~230
2.7~30
2.7~22.0
10~210
1.3~28
1.3~20.0
88
Infant
Neonate
40~200
40~135
5.3~18.0
20~165
20~110
2.7~14.6
10~150
10~95
1.3~12.7