Step 1: rotate the knob to move the gray cursor to the setting item and press the knob to confirm your selection.
Step 2: rotate the knob to change the setting or modify the setting value.
Step 3: press the knob again to change and repress it to save the setting.
Pressing "
" key to return to the Main Screen.
Long pressing Navigation Knob to enter into "System Menu" screen.
5.1 System Parameter Settings
On "System Menu" screen, select "Setup"→"SYS" to enter into system related settings.
Type: The category of patient being monitored, this can be selected from Adult, Pediatric and Neonate.
Adult: the patient is adult.
Pediatric: the patient is pediatric.
Neonate: the patient is neonate.
The default is "Adult"
When changing the patient type, the monitor will change the default alarm settings, and initialize the
NIBP and SpO
starting the monitoring. It is forbidden to use Adult type on the Pediatric patient, or it may cause serious
injury.
Mode: Monitor working mode selection. The "Real Time" shows the normal working mode with real time
waveform and data. The "Demo" shows the demo working mode with demo waveform and data. In the demo
mode, all the signals and data are generated by the device for demonstrating and testing purpose. The default is
"Real Time"
Operator's Manual for Patient Monitor
Chapter 5 Parameter Settings
Figure 5.1 System Setup
measuring module accordingly. Please pay special attention to the patient type before