13.1 Overview

The monitor has built-in chargeable battery, when the monitor is disconnected from the external AC power supply,
then it turns to be powered by built-in battery. When the monitor is supplied by external AC power, then the battery
will be charged. Generally, it's unnecessary to maintain the battery when using under normal working condition.
When the monitor is powered by built-in battery, and battery is in low voltage, then technical alarm will be triggered,
and message "Low battery" pops up. At this moment, the user should connect the external power immediately to
make sure the monitor continue working.
The capacity of the internal battery is limited. If the battery charge is too low, a technical alarm will be triggered and
the message displayed. At this moment, apply AC/DC power to the monitor. Otherwise, the monitor will power off
automatically before the battery is completely depleted.

13.2 Battery Maintenance

Please pay attention to the polarity of battery, do NOT insert it into battery compartment with reversed
polarities;
Do NOT use the batteries manufactured by other companies, if being inserted, the device will may be
damaged;
In order to avoid damaging the battery, do NOT use other power supply device to charge the battery;
After battery ageing phenomenon occurs, do NOT throw the battery into fire to avoid explosion risk.
Do not hit or strike it with force;
Do not use this battery on other devices;
Do not use this battery below -10°C or above 40°C;
Dispose of the battery, the local law should be followed.
In order to maintain battery supply time and prolong battery lifetime, please charge the battery every one or
two months if don't use battery for a long time. Charge battery at least 12-15 hours every time. Before
charging, run down the internal battery until the monitor turn off automatically to minimize memory effects.
Charging time will be the same no matter whether the monitor is working or not. Charge fully before putting
the monitor into storage.
Using a monitor powered solely by an internal battery which has short charge power will cause the monitor
turn off automatically when the battery is depleted..
Do not use batteries manufactured by other companies, which may cause damage to the device. If battery is
damaged, please replace with same type and specification battery marked by "CCC" or "CE" in time, or contact
the company directly.