2.3.1 Information Area
The Information Area will display the patient information (including gender, patient ID, patient type and name),
physiological alarm sources, alarm sound status and current date.
Alarm status
Patient type
"Alarm
": Audible alarm status, green "
alarm sound is silent. The audible alarm will be activated again automatically after 2 minutes or when a new
alarm event occurs. Red " " indicates the audible alarm is disabled, that means the alarm sound is off, this is also
normal situation when the alarm sound volume is set to "0" in system parameter settings.
"ADUL": The patient type. There are two types available: "Adult" and "Pediatric".
"MON": ECG Filter type. There are three types: "Diagnosis", "Monitor" and "Operation". The option can be set
in the setting menu.
: network connection indicator. "
monitoring system yet; "
successfully.
: battery voltage indicator; When the indicator is yellow and displays only one "grid", it means there is
a little battery power left. When the indicator turns red and blinks, as well as less than one "grid" displays, the
system alarm will be on to remind the low battery voltage. Please connect the device to the mains power supply
in time to ensure the normal use of monitor, and the battery will be recharged. When the battery is fully charged,
battery voltage indicator displays full grid. During recharging, the grids in the battery indicator are rolling
circularly.
"2011-03-02 13:57:25": the current calendar time and date. The system time and date can be set during the
system start-up when the screen displays the time and data setups. The current figure shows the time and date is
nd
March 2
, 13:57:25, 2011.
"Long press knob to enter System Menu": System prompt or description for the current status.
"PID": The patient ID. The patient ID can be entered or changed in the archive management window
2.3.2 Parameter Area and Waveform Area
1) Parameter Area
Parameter area displays each parameter's value, unit and icon etc.. Move Navigation Knob to focus a certain
parameter panel, and the panel (such as ECG parameter panel shown in below figure) will be highlighted, then press
Knob can enter into the corresponding parameter setting screen.
Operator's Manual for Patient Monitor
ECG Filter type
" indicates the audible alarm is enabled, yellow " " indicates the
" means the patient monitor is not linked with the central
" (in green) means the patient monitor is linked to the central monitoring system
14
Network connection indicator
Date and Time