7 Faults

7
Faults
7.1 Faults when using with
Fault
The
symbol does not appear on
the therapy device display.
The
symbol appears in the
therapy device display, but the analog
values are "0" or not plausible.

7.2 Faults when using with prisma HUB

36
EN EN prismaPSG
If you are not able to clear an error message with the aid of the
table below, the device functions differently than expected, or an
event occurs, you should have the device repaired by your
authorized dealer. To avoid serious damage, do not continue using
the device.
prismaCONNECT
Cause
The PSG connection cable is not
inserted or is defective.
Faulty or non-existent connection
to PSG.
Please observe the instructions for use for the prisma HUB module.
Remedy
Disconnect and reconnect the
PSG connection cable.
Check the connections to the
PSG.