Removing The Prismapsg Module; Scaling On The Psg System - Lowenstein Medical PrismaPSG Mode D'emploi

Module pour appareils de thérapie du type wm 090 td, wm 100 td, wm 110 td, wm 120 td
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

4 Preparation and operation

4.2 Removing the prismaPSG module

Result

4.3 Scaling on the PSG system

32
EN
prismaPSG
1. Remove the connector of the PSG connection cable from the
module interface on the prismaCONNECT or prisma HUB
module.
The module is removed.
The prismaPSG module has 8 analog connection sockets. These are
assigned set parameters:
Analog
output
Signal
socket
1
Resp. flow
2
Unintentional leak
3
Obstruction level
4
Setpoint pressure
5
Actual pressure
6
Minute volume
7
Tidal volume
8
Frequency
To label the assignment of the channels, cut out one of the strips
of paper on the last page of the instructions for use off and insert
it in the front panel membrane of the module prismaPSG.
Lower phys.
Upper phys.
limit
limit
-250 l/min
0 l/min
0 %
0 hPa
0 hPa
0 l/min
0 ml
0 bpm
250 l/min
100 l/min
100 %
40 hPa
40 hPa
63 l/min
2000 ml
63 bpm