Starting The Sterilization Cycle; End Of Cycle; Discharging Used Water - Tecno-gaz Europa B evo Mode D'emploi

L'autoclave
Masquer les pouces Voir aussi pour Europa B evo:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

8.6

Starting the sterilization cycle

Choose sterilization program most adaptable of load prepared and press the button
B-SELECT.
Each individual pressure changes program selection to the next.
After having chosen the program start the cycle by pressing the button B-START, the door
will lock automatically and the cycle will begin.
During the cycle, the display will show all parameters and relative information.
8.7

End of cycle

An acoustic signal will warn the operators that the sterilization cycle has been completed.
If the cycle is concluded with a positive outcome, the message END OF CYCLE will appear
on the display; if, on the other hand, problems arise, an error or alarm code will be
displayed.
When the acoustic signal stops, release the door by pressing the
is pressure inside the chamber, the button will not activate the release device. Wait until the
chamber is completely depressurized, then repeat the operation or, in emergency situations
only, simultaneously press
handle to open it.
8.8
Unloading, preserving and storing the sterilizied materials
Wear personal protection equipment in accordance with the existing laws.
Extract the trays using the special spanner provided (PICT.10), and place the autoclave
load on a clean, dry surface.
Then store the treated instruments in the special, prepared compartments.
8.9

Discharging used water

When the led of used water level (PICT.2 pos. 28.) goes on, proceed with emptying. If not
emptied, the autoclave function is inhibited.
Take the pipe provided (PICT.16), and insert it in the used water discharge plug, on the
front of the autoclave (PICT.2 pos. 14.), unscrew the metal ring anticlockwise (to execute 2
complete turns), simultaneously place the other end of the pipe in a container, the water will
descend into the same container.
IMPORTANT:
A – The hose located in the receiving container, must not ever, touch or be immersed in the
discharged water, otherwise there will be a situation of sucking up.
B – Always wait until the unloading water is totally emptied. The led of maximum used
water level will turn off when there is still water in the tank, therefore do not use as a
reference for this operation.
To the term of the drainage to screw the metal ring and remove the tube.
+ B-PUMP. When the door is released, pull the door
B-DOOR
ENGLISH
button. If there
B-DOOR
23

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Europa b evo 24

Table des Matières