Messages of error are emphasized through an Alfa-numerical code, consisting in a letter
and three numbers.
The "E" CODE is relative to errors of the operator, which can be corrected by same
operator.
The "A" CODE is relative to alarms, irregularities of the autoclave, after having carried out
the remedy, if the problem persists it is necessary to request telephonic assistance.
If an alarm message is displayed (code "A") the cycle IS TO BE
CONSIDERED ABORTED: it will be necessary to repeat all the preparation
and sterilization operations
To reset alarms and errors, maintain pressed the buttons here indicated until the
intervention of a
BEEP"
"
B-DOOR
10.1 List of error alarm messages
ERROR
Low water level
E 200
Low level water cleaned up with loadingfrom
E 250
demineralizer
Maximum level used water
E 300
Door open
E 401
Door unlocked
E 451
High temperature in chamber for vacuum test
E 461
(above 40° C)
Cycle begins without water
A 250
Interrupted cycle
A 001
Irregular pc display
A 011
Vacuum not reached in 10' (minutes)
A 101
Vacuum not maintened for first phase VT
A 111
Vaccum not maintened for second phase VT
A 121
Hatch opende during the cycle or problem
A 401
closing
28
MESSAGES OF ERROR OR ALARMS
+
B-PUMP
CAUSE
+
B-SELECT
REMEDY
• Fill tank
• Fill tank
• Empty tank
• Colse door with attention
• Open door and close
again
• Shut off and wait 10
minutes
• Control demineralizer
device
• Reset and restart cycle
• Shut off and restart
autoclave
• Reset and restart cycle
• Reset and restart cycle
• Reset and restart cycle
• Check the closing system