Pressure not maintened
A 501
Irregular 1° temperature sensor
A 551
Irregular 2° temperature sensor
A 552
Overly high pressure
A 641
Overlay high temperature
A 651
Irregular temperature comparison
A 661
Working pressure not reached
A 701
Low temperature
A 751
Irregular pressor sensor
A 761
Irregular unloading pressure
A 801
Main voltage too low - Temporary lack of
A 901
electric energy
REPLACE RESINS Resins IONIC
SYSTEM exhaust
REPLACE FILTER Resins OSMOSI
SYSTEM exhaust
The printer's door is not
D
closed
Paper is missing
P
There is low voltage on the
Lo
line
In the event of one of the alarms persisting after some time, consult technical assistance.
Replacement of resins; refer to paragraph
14.3
Replacement of filter; refer to paragraph 14.3
Try to close the printer's door
Try to connect the machine to the net electrical worker in
an other point; if the problem persists to contact a
• Reset and restart cycle
• Reset and restart cycle
• Reset and restart cycle
• Reset and restart cycle
• Reset and restart cycle
• Reset and restart cycle
• Reset and restart cycle
• Reset and restart cycle
• Reset and restart cycle
• Reset and restart cycle
• Reset and restart cycle,
• Inform eletric Energy
provider
Insert paper
electrician
ENGLISH